Psurio
Post-market surveillance snapshot

Volumetric infusion pump (sample)

Controlled intravenous delivery of fluids and medicines to adults and children in hospital.
EU class IIb · FDA product code FRN
Surveillance period 15 October 2025 to 15 October 2026
Draft: 80 literature records still awaiting review
53searches run and logged
168unique literature records
0records included
51,335adverse-event reports
24FDA recalls
15signals to investigate

Prepared 15 October 2026 · Snapshot 2026-10-15-volumetric-infusion-pump-sample-858765

Prepared with Psurio

1. Signals worth investigating

PrioritySignalEvidence
highRising: Pumping Problem179 reports in 2026-07-03 to 2026-09-30 (60.54/month), 15.27× the earlier monthly rate.
highRising: Moisture Damage49 reports in 2026-07-03 to 2026-09-30 (16.57/month), 71.05× the earlier monthly rate.
highRising: Device Displays Incorrect Message48 reports in 2026-07-03 to 2026-09-30 (16.23/month), 6.33× the earlier monthly rate.
highRising: Device Reprocessing Problem44 reports in 2026-07-03 to 2026-09-30 (14.88/month), new in the recent window.
mediumRising: Failure to Infuse1109 reports in 2026-07-03 to 2026-09-30 (375.09/month), 2.03× the earlier monthly rate.
mediumANSM field safety notice: Pompe à perfusion ambulatoire (accessoire) – Cassette-réservoirs CADD avec raccord NRFit – ICU Medical Inc.2026-09-16: Rappel de produit · Rappel n° R2625748 destiné aux pharmacies d'usage intérieur et aux prestataires de santé à domicile (matched device term: pompe à perfusion).
mediumMHRA field safety notice: ICU Medical: CADD Medication Cassette Reservoirs with NRFit2026-08-28: Infusion pump cassette (matched device term: infusion pump).
mediumBfArM field safety notice: Urgent Field Safety Notice for AlarisTM GW 800 Volumetric Pump by BD Switzerland Sarl2026-08-05 (matched device term: volumetric pump).
mediumANSM field safety notice: Pompe à perfusion Alaris GW 800 – BD2026-08-04: Information aux utilisateurs · Information n° R2622916 destinée aux pharmacies d'usage intérieur, établissements de santé (matched device term: pompe à perfusion).
mediumSwissmedic field safety notice: MEDCAPTAIN LIFE SCIENCE CO., LTD.: BD neXus Disposable Infusion Set2026-07-22: Electric infusion pump administration set, single-use (update) (matched device term: infusion pump).
mediumFDA recall by Baxter Healthcare Corporation (2 products)Initiated 2026-08-11: A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.
mediumFDA recall by ICU Medical, Inc. (2 products)Initiated 2026-06-10: In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.
mediumFDA recall by ICU Medical, Inc.Initiated 2026-05-12: Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and…
mediumFDA recall by Fresenius Kabi USA, LLCInitiated 2026-05-06: Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
mediumFDA recall by Fresenius Kabi USA, LLCInitiated 2026-05-06: Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service.…
info93 reports coded as deathMAUDE reports don't establish that the device caused the death. Read each narrative.
info16 more field safety notices in the periodListed in full in the field safety notices section.
info13 more recall events in the periodListed in full in the recalls table.

A signal is a prompt to investigate, not a conclusion. Thresholds: a monthly spike above the mean of earlier months plus two standard deviations; a problem reported at least twice as often per month in the last 90 days, with at least 5 reports; a complaint category significantly above its earlier rate (one-sided Poisson test, p < 0.01, at least 3 complaints).

2. Adverse events (FDA MAUDE)

Reports per month

0 2,500 5,000 7,500 10,000Oct 25Dec 25Feb 26Apr 26Jun 26Aug 26Sept 26
Volumetric infusion pump (sample) (51,335 reports)

Hollow points are incomplete months. MAUDE data for this device runs to 2026-09-30. Counts are not adjusted for the number of devices in use.

Event types

TypeReports
Malfunction50,321
Injury673
Other248
Death93

Device problems: last 90 days versus before

Recent: 2026-07-03 to 2026-09-30. Earlier: 2025-10-15 to 2026-07-02. Rates are per month. Rows marked “Exact count” were counted one problem at a time; rows marked “Estimate” come from the FDA’s list of the most frequent problems and may undercount less common ones.

ProblemRecentEarlierRecent / monthEarlier / monthChangeCount
Failure to Infuse1,1091,582375.1184.52.03×Exact count
Pumping Problem1793460.5415.27×Exact count
Moisture Damage49216.60.271.05×Exact count
Device Displays Incorrect Message482216.22.66.33×Exact count
Device Reprocessing Problem44014.90newExact count
Material Discolored44014.90newExact count
Leak/Splash423514.24.13.48×Exact count
Inaccurate Flow Rate344111.54.82.4×Exact count
Defective Device24318.13.62.25×Exact count
Intermittent Infusion23147.81.64.76×Exact count
Device Alarm System2,1293,649720.1425.61.69×Estimate
Device Sensing Problem1,7875,882604.46860.88×Estimate
Break1,3264,125448.5481.10.93×Estimate
Failure to Sense1,3153,875444.8451.90.98×Estimate
No Apparent Adverse Event1,1284,053381.5472.70.81×Estimate

Most reported patient problems

Patient problemReports
No Clinical Signs, Symptoms or Conditions48660
Insufficient Information1110
Hyperglycemia944
Low Blood Pressure/ Hypotension153
Hypoglycemia87
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available85
High Blood Pressure/ Hypertension50
Tachycardia39
Hemorrhage/Blood Loss/Bleeding36
Reaction to Medicinal Component of Device32

FDA recalls in the period

24 recalled products in 18 recall events, newest first.

InitiatedFirmReasonRecall numbersStatus
2026-08-11Baxter Healthcare CorporationA software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains.Z-3065-2026
Z-3066-2026
Open, Classified
2026-06-10ICU Medical, Inc.In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.Z-2723-2026
Z-2724-2026
Open, Classified
2026-05-12ICU Medical, Inc.Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out.Z-3162-2026Open, Classified
2026-05-06Fresenius Kabi USA, LLCPotential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.Z-2223-2026Open, Classified
2026-05-06Fresenius Kabi USA, LLCFresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern.Z-2224-2026Open, Classified
2026-04-29ICU Medical, Inc.Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery.Z-2311-2026
Z-2312-2026
Z-2313-2026
Open, Classified
2026-02-11CareFusion 303, Inc.Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care UnitsZ-1645-2026Open, Classified
2026-02-04Fresenius Kabi USA, LLCPotential for the Cassette Loading Lever to break.Z-2274-2026Open, Classified
2026-01-13Baxter Healthcare CorporationPumps were released without full testing being performed, including occlusion alarm testing.Z-1368-2026Open, Classified
2025-12-15ICU Medical, Inc.ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or provide audible feedback for touchscreen inputs.Z-1094-2026Open, Classified
2025-12-15ICU Medical, Inc.Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed.Z-1118-2026
Z-1119-2026
Z-1120-2026
Open, Classified
2025-11-28Baxter Healthcare CorporationCertain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation.Z-1121-2026Open, Classified
2025-11-21Fresenius Kabi USA, LLCEmphasizing instructions for LVP duration programming located in the IFU.Z-1048-2026Open, Classified
2025-11-14Fresenius Kabi USA, LLCSoftware version 5.10.1 and earlier contain anomalies that have the potential to cause serious patient harm or death.Z-0885-2026Open, Classified
2025-11-06CareFusion 303, Inc.Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital networkZ-1217-2026Open, Classified
2025-11-03Fresenius Kabi USA, LLCIncorrect assembly of specific lot of LVP Primary Administration Set that could result in serious patient injury.Z-0586-2026Open, Classified
2025-11-03Fresenius Kabi USA, LLCDownstream Occlusion alarms may occur during procedures using low flow infusion rates when a check valve* is added to the Ivenix Administration Set.Z-0848-2026Open, Classified
2025-10-17CareFusion 303, Inc.If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.Z-0430-2026Open, Classified

3. Field safety notices (UK, Germany, France, Switzerland, Spain)

34 notices from European regulators (MHRA, BfArM, ANSM, Swissmedic, AEMPS) in the period, of which 13 only mention a search term inside the notice document.

DateSourceNoticeMatched on
2026-09-16ANSM
Field safety notice
Pompe à perfusion ambulatoire (accessoire) – Cassette-réservoirs CADD avec raccord NRFit – ICU Medical Inc.
Rappel de produit · Rappel n° R2625748 destiné aux pharmacies d'usage intérieur et aux prestataires de santé à domicile
device term: pompe à perfusion
2026-08-28MHRA
Field safety notice
ICU Medical: CADD Medication Cassette Reservoirs with NRFit
Infusion pump cassette
Ref 40630462 2026/008/025/601/011
device term: infusion pump
2026-08-05BfArM
Field safety notice
Urgent Field Safety Notice for AlarisTM GW 800 Volumetric Pump by BD Switzerland Sarldevice term: volumetric pump
2026-08-04ANSM
Field safety notice
Pompe à perfusion Alaris GW 800 – BD
Information aux utilisateurs · Information n° R2622916 destinée aux pharmacies d'usage intérieur, établissements de santé
device term: pompe à perfusion
2026-07-22Swissmedic
Field safety notice
MEDCAPTAIN LIFE SCIENCE CO., LTD.: BD neXus Disposable Infusion Set
Electric infusion pump administration set, single-use (update)
Ref Vk_20260713_032
device term: infusion pump
2026-07-02Swissmedic
Field safety notice
B. Braun Melsungen AG: Infusomat Space P
Bedside infusion pump, single-channel
Ref Vk_20260701_035
device term: infusion pump
2026-03-31MHRA
Field safety notice
Abiomed, Inc: Impella Purge Cassette
Infusion Pump Cassette · Model: 0043-0001;0043-0009
Ref 39420703 2026/004/002/601/034
device term: infusion pump
2026-02-16ANSM
Field safety notice
Pompe à perfusion ambulatoire (accessoires) – Set d’administration CADD à débit élevé – ICU Medical Inc.
Rappel de produit · Rappel n° R2604556 destiné aux pharmacies d'usage intérieur des établissements de santé et aux prestataires de santé à domicile
device term: pompe à perfusion
2026-02-13MHRA
Field safety notice
ICU Medical: CADD High Volume Administration Sets
Electric infusion pump administration set · Model: 21-7361-24
Ref 38755259 2026/002/010/601/028
device term: infusion pump
2026-02-12Swissmedic
Field safety notice
ICU Medical Inc: CADD™ High Volume Administration Sets
Electric infusion pump administration set, single-use
Ref Vk_20260210_023
device term: infusion pump
2026-02-05ANSM
Field safety notice
Pompe à perfusion (tubulure) – Sets de perfusion Alaris – BD
Rappel de produit · Information n° R2602319 destinée aux pharmacies d'usage intérieur, établissements de santé
device term: pompe à perfusion
2026-01-28MHRA
Field safety notice
Baxter: EVO IQ Large Volume Pump
Model: ELVP002UKI and 04400121200
Ref 38329736 2026/001/019/601/021
device term: large volume pump
2026-01-27Swissmedic
Field safety notice
BD Switzerland Sàrl: Alaris VP Infusion Sets
Electric infusion pump administration set, single-use
Ref Vk_20260122_01
device term: infusion pump
2026-01-21MHRA
Field safety notice
BD: Alaris VP Infusion Sets
Infusion Pump Administration Set
Ref 38309876 2026/001/014/601/056
device term: infusion pump
2025-12-17ANSM
Field safety notice
Pompe à perfusion BD neXus V700 et pousse-seringue – BD neXus S700 – Medcaptain Life Science Co.
Information aux utilisateurs · Information n° R2534072 destinée aux pharmacies d'usage intérieur des établissements de santé
device term: pompe à perfusion
2025-12-16BfArM
Field safety notice
Urgent Field Safety Notice for BD Alaris System Infusion Pump by BD Switzerland Sarldevice term: infusion pump
2025-12-15MHRA
Field safety notice
MEDCAPTAIN MEDICAL TECHNOLOGY: BD neXus V700 Infusion Pump & S700 Syringe Pump
V700: Infusion Pump; S700: Syringe Pump
Ref 37923637 2025/012/008/601/002
device term: infusion pump
2025-12-01MHRA
Field safety notice
Baxter: Ambulatory Elastomeric Infusion Pumps
Ref 37850863 2025/012/001/601/064
device term: infusion pump
2025-12-01BfArM
Field safety notice
Urgent Field Safety Notice for Ambulatory Elastomeric Infusion Pumps by Baxter Healthcare S. (Opfikon)device term: infusion pump
2025-10-28MHRA
Field safety notice
Baxter Healthcare SA: EVO IQ Large Volume Pump
15th August 2025 Bedside infusion pump, single-channel MHRA reference: 37397515 2024/011/028/601/018 – Customer reply form · Model: ELVP001UKI
Ref 37397515 2024/011/028/601/018
device term: infusion pump
2025-10-16MHRA
Field safety notice
ICU Medical: CADD-Solis Ambulatory Infusion Pumps
Bedside infusion pump, single-channel · Model: CADD-Solis VIP
Ref 37337852 2025/010/013/601/015
device term: infusion pump

Mentioned only inside the notice document (BfArM)

BfArM's search matched these in the notice text. Check whether each concerns the device.

DateSourceNoticeMatched on
2026-09-02BfArM
Field safety notice
Urgent Field Safety Notice for IntelliSpace Critical Care and Anesthesia by Philips Healthcaredevice term: infusion pump (in document text)
2026-08-19BfArM
Field safety notice
Urgent Field Safety Notice for ORBIS Medication by DH Healthcare GmbHdevice term: infusion pump (in document text)
2026-07-20BfArM
Field safety notice
Urgent Field Safety Notice for Infusomat Space P by B. Braun Melsungen AG , Sparte Hospital Caredevice term: infusion pump (in document text)
2026-07-17BfArM
Field safety notice
Urgent Field Safety Notice for BD Alaris Plus and Alaris neXus Syringe Pumps by BD Switzerland Sarldevice term: infusion pump (in document text)
2026-07-16BfArM
Field safety notice
Urgent Field Safety Notice for Introducer Kit for Impella by Oscor, Incdevice term: smart pump (in document text)
2026-07-16BfArM
Field safety notice
Urgent Field Safety Notice for Introducer Kit for Impella by Oscor, Incdevice term: smart pump (in document text)
2026-07-15BfArM
Field safety notice
Urgent Field Safety Notice for Disposable Infusion Set by MEDCAPTAIN MEDICAL TECHNOLOGY CO ., LTD.device term: infusion pump (in document text)
2026-05-07BfArM
Field safety notice
Urgent Field Safety Notice for Spaceplus Perfusor by B. Braun Melsungen AGdevice term: infusion pump (in document text)
2026-04-20BfArM
Field safety notice
Urgent Field Safety Notice for Kaleido by ViCentra B.V.device term: infusion pump (in document text)
2026-04-13BfArM
Field safety notice
Urgent Field Safety Notice for SynchroMed II Clinician Programmer Application by Medtronic Incdevice term: infusion pump (in document text)
2026-02-26BfArM
Field safety notice
Urgent Field Safety Notice for MyPTM App by Medtronic Incdevice term: infusion pump (in document text)
2026-01-23BfArM
Field safety notice
Urgent Field Safety Notice for AlarisTM VP Infusion Sets by BD Switzerland Sarldevice term: infusion pump (in document text)
2025-10-30BfArM
Field safety notice
Urgent Field Safety Notice for CamAPS FX App und CamAPS and mylife CamAPS by CamDiab Ltddevice term: smart pump (in document text)

4. Literature

Screening flow

Records found: 190
↓ 22 duplicates removed
Unique records: 168
↓ 88 excluded by rules
Decided by a person: 0 (0 excluded)
↓
Included: 0
80 still awaiting review

Decisions by relevance band

Records were ordered for review by a transparent relevance score (device in title or subject indexing, mentions in the abstract, safety topics). The score only ordered the work; every record was decided individually.

BandRecordsIncludedExcludedAwaiting
high260026
medium150015
low12708839

Exclusions by reason

CodeReasonRecords
X3No safety or performance data (editorial, news)1
X1Not the device type under surveillance87

Included records and appraisal

No records included yet. Review the queue to complete this section.

5. Latest adverse-event reports

ReceivedReport
2026-09-30
Malfunction
Spectrum IQ BAXTER HEALTHCARE CORPORATION · 1314492-2026-02104
False Alarm
2026-09-30
Malfunction
Alaris System CAREFUSION 303, INC. · 2016493-2026-67346
Break; Contamination
It was reported that the device was missing a piece. There was no patient involvement.
2026-09-30
Malfunction
Alaris System CAREFUSION 303, INC. · 2016493-2026-67349
Circuit Failure; Physical Resistance/Sticking
It was reported that the device had pump will not power on. There was no patient involvement.
2026-09-30
Malfunction
Alaris System CAREFUSION 303, INC. · 2016493-2026-67350
Contamination
It was reported that the device had not turning on, replaced battery, same issue. There was no patient involvement.
2026-09-30
Malfunction
Alaris System CAREFUSION 303, INC. · 2016493-2026-67354
Device Sensing Problem
It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.
2026-09-30
Malfunction
Alaris System CAREFUSION 303, INC. · 2016493-2026-67353
Calibration Problem
It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Field Service]";"[Door ajar-safety clamp test failed]". There was no reported patient involvement.
2026-09-30
Malfunction
Alaris System CAREFUSION 303, INC. · 2016493-2026-67347
Device Alarm System
It was reported that the device had error code 242.4030. There was no patient involvement.
2026-09-30
Malfunction
Alaris System CAREFUSION 303, INC. · 2016493-2026-67351
Circuit Failure
It was reported that the device washed out screen. There was patient involvement but unknown harm.
2026-09-30
Malfunction
Alaris System CAREFUSION 303, INC. · 2016493-2026-67348
Failure to Sense
It was reported that the device displayed channel error code 242.4030. There was no patient involvement.
2026-09-30
Malfunction
Alaris System CAREFUSION 303, INC. · 2016493-2026-67355
Premature Discharge of Battery
It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.
2026-09-30
Malfunction
IVENIX INFUSION SYSTEM FRESENIUS KABI USA, LLC · 3014732157-2026-04133
No Apparent Adverse Event
The following has been reported:The unit was returned during service after initial complaint investigation was completed for rear housing issues. Secondary Findings: Unit experienced tubing set misloaded warnings due to sticky FVA pins. Pins and channels were cleaned and unit then passed a mock infusion.
2026-09-30
Malfunction
IVENIX INFUSION SYSTEM FRESENIUS KABI USA, LLC · 3014732157-2026-04134
No Apparent Adverse Event
The following has been reported:The unit was returned during service after initial complaint investigation was completed for lever arm issues. Secondary Findings: Secondary Findings: Outlet FVA pin stuck in due to excess debris. FVA pin lip seals will be replaced during repair.

6. Limitations

Appendix A. Searches run

SourceQueryHitsRetrievedRun at (UTC)
PubMed("infusion pump"[tiab] OR "volumetric pump"[tiab] OR "smart pump"[tiab] OR "large volume pump"[tiab] OR "Infusion Pumps"[mh]) AND ("adverse event"[tiab] OR "malfunction"[tiab] OR "device failure"[tiab] OR "equipment failure"[tiab] OR "recall"[tiab] OR "occlusion"[tiab] OR "alarm"[tiab] OR "free flow"[tiab] OR "over-infusion"[tiab] OR "under-infusion"[tiab] OR "medication error"[tiab] OR "programming error"[tiab] OR "dose error"[tiab] OR "drug library"[tiab] OR "use error"[tiab] OR "interoperability"[tiab] OR "Equipment Failure"[mh] OR "Equipment Safety"[mh] OR "Medication Errors"[mh] OR "Equipment Failure Analysis"[mh] OR "adverse effects"[sh]) | publication date 2025/10/15 to 2026/10/151671672026-10-15 18:22:46
Europe PMC(TITLE_ABS:"infusion pump" OR TITLE_ABS:"volumetric pump" OR TITLE_ABS:"smart pump" OR TITLE_ABS:"large volume pump") AND (TITLE_ABS:"adverse event" OR TITLE_ABS:"malfunction" OR TITLE_ABS:"device failure" OR TITLE_ABS:"equipment failure" OR TITLE_ABS:"recall" OR TITLE_ABS:"occlusion" OR TITLE_ABS:"alarm" OR TITLE_ABS:"free flow" OR TITLE_ABS:"over-infusion" OR TITLE_ABS:"under-infusion" OR TITLE_ABS:"medication error" OR TITLE_ABS:"programming error" OR TITLE_ABS:"dose error" OR TITLE_ABS:"drug library" OR TITLE_ABS:"use error" OR TITLE_ABS:"interoperability") AND FIRST_PDATE:[2025-10-15 TO 2026-10-15] AND (SRC:MED OR SRC:PMC OR SRC:PPR)23232026-10-15 18:22:46
openFDA MAUDE(device.device_report_product_code:"FRN") AND date_received:[20251015 TO 20261015] | count by date_received513353122026-10-15 18:22:47
openFDA MAUDE(device.device_report_product_code:"FRN") AND date_received:[20260703 TO 20260930] | count by product_problems.exact166231002026-10-15 18:22:47
openFDA MAUDE(device.device_report_product_code:"FRN") AND date_received:[20251015 TO 20260702] | count by product_problems.exact493941002026-10-15 18:22:47
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Failure to Infuse" AND date_received:[20260703 TO 20260930]110912026-10-15 18:22:48
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Failure to Infuse" AND date_received:[20251015 TO 20260702]158212026-10-15 18:22:48
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Pumping Problem" AND date_received:[20260703 TO 20260930]17912026-10-15 18:22:49
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Pumping Problem" AND date_received:[20251015 TO 20260702]3412026-10-15 18:22:49
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Moisture Damage" AND date_received:[20260703 TO 20260930]4912026-10-15 18:22:49
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Moisture Damage" AND date_received:[20251015 TO 20260702]212026-10-15 18:22:49
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Device Displays Incorrect Message" AND date_received:[20260703 TO 20260930]4812026-10-15 18:22:50
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Device Displays Incorrect Message" AND date_received:[20251015 TO 20260702]2212026-10-15 18:22:50
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Device Reprocessing Problem" AND date_received:[20260703 TO 20260930]4412026-10-15 18:22:50
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Device Reprocessing Problem" AND date_received:[20251015 TO 20260702]002026-10-15 18:22:51
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Material Discolored" AND date_received:[20260703 TO 20260930]4412026-10-15 18:22:51
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Material Discolored" AND date_received:[20251015 TO 20260702]002026-10-15 18:22:51
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Leak/Splash" AND date_received:[20260703 TO 20260930]4212026-10-15 18:22:52
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Leak/Splash" AND date_received:[20251015 TO 20260702]3512026-10-15 18:22:52
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Inaccurate Flow Rate" AND date_received:[20260703 TO 20260930]3412026-10-15 18:22:52
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Inaccurate Flow Rate" AND date_received:[20251015 TO 20260702]4112026-10-15 18:22:53
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Defective Device" AND date_received:[20260703 TO 20260930]2412026-10-15 18:22:53
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Defective Device" AND date_received:[20251015 TO 20260702]3112026-10-15 18:22:53
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Intermittent Infusion" AND date_received:[20260703 TO 20260930]2312026-10-15 18:22:53
openFDA MAUDE(device.device_report_product_code:"FRN") AND product_problems.exact:"Intermittent Infusion" AND date_received:[20251015 TO 20260702]1412026-10-15 18:22:54
openFDA MAUDE(device.device_report_product_code:"FRN") AND date_received:[20251015 TO 20261015] | count by event_type.exact5133542026-10-15 18:22:54
openFDA MAUDE(device.device_report_product_code:"FRN") AND date_received:[20251015 TO 20261015] | count by patient.patient_problems.exact51500252026-10-15 18:22:54
openFDA MAUDE(device.device_report_product_code:"FRN") AND date_received:[20251015 TO 20261015] | sorted by date_received:desc513351002026-10-15 18:22:55
openFDA recalls(product_code:"FRN") AND event_date_initiated:[2025-10-15 TO 2026-10-15]24242026-10-15 18:22:57
MHRAGOV.UK medical safety alerts of type field-safety-notices (and titled "Field Safety Notices"), published 2025-10-15 to 2026-10-15; each notice matched against: infusion pump, volumetric pump, smart pump, large volume pump, pompe à perfusion, pompes à perfusion, Infusionspumpe, pompa infusionale, pompa per infusione, bomba de infusión55392026-10-15 18:23:04
MHRAGOV.UK medical safety alerts of type device-safety-information, published 2025-10-15 to 2026-10-15; title and summary matched against the same terms1402026-10-15 18:23:04
BfArMField safety notices (Kundeninfo) containing "infusion pump", newest first, dated 2025-10-15 to 2026-10-1514142026-10-15 18:23:06
BfArMField safety notices (Kundeninfo) containing "volumetric pump", newest first, dated 2025-10-15 to 2026-10-15332026-10-15 18:23:08
BfArMField safety notices (Kundeninfo) containing "smart pump", newest first, dated 2025-10-15 to 2026-10-15442026-10-15 18:23:10
BfArMField safety notices (Kundeninfo) containing "large volume pump", newest first, dated 2025-10-15 to 2026-10-15002026-10-15 18:23:13
BfArMField safety notices (Kundeninfo) containing "pompe à perfusion", newest first, dated 2025-10-15 to 2026-10-15002026-10-15 18:23:14
BfArMField safety notices (Kundeninfo) containing "pompes à perfusion", newest first, dated 2025-10-15 to 2026-10-15002026-10-15 18:23:16
BfArMField safety notices (Kundeninfo) containing "Infusionspumpe", newest first, dated 2025-10-15 to 2026-10-15002026-10-15 18:23:17
BfArMField safety notices (Kundeninfo) containing "pompa infusionale", newest first, dated 2025-10-15 to 2026-10-15002026-10-15 18:23:19
BfArMField safety notices (Kundeninfo) containing "pompa per infusione", newest first, dated 2025-10-15 to 2026-10-15002026-10-15 18:23:20
BfArMField safety notices (Kundeninfo) containing "bomba de infusión", newest first, dated 2025-10-15 to 2026-10-15002026-10-15 18:23:21
ANSMInformations de sécurité, dispositifs médicaux, 2025-10-15 to 2026-10-15; titles matched against 10 terms5695692026-10-15 18:23:36
SwissmedicFSCA publications matching "infusion pump", 2025-10-15 to 2026-10-1515152026-10-15 18:23:36
SwissmedicFSCA publications matching "volumetric pump", 2025-10-15 to 2026-10-15112026-10-15 18:23:36
SwissmedicFSCA publications matching "smart pump", 2025-10-15 to 2026-10-15222026-10-15 18:23:37
SwissmedicFSCA publications matching "large volume pump", 2025-10-15 to 2026-10-15112026-10-15 18:23:37
SwissmedicFSCA publications matching "pompe à perfusion", 2025-10-15 to 2026-10-15882026-10-15 18:23:37
SwissmedicFSCA publications matching "pompes à perfusion", 2025-10-15 to 2026-10-15442026-10-15 18:23:38
SwissmedicFSCA publications matching "Infusionspumpe", 2025-10-15 to 2026-10-15332026-10-15 18:23:38
SwissmedicFSCA publications matching "pompa infusionale", 2025-10-15 to 2026-10-15112026-10-15 18:23:39
SwissmedicFSCA publications matching "pompa per infusione", 2025-10-15 to 2026-10-15552026-10-15 18:23:39
SwissmedicFSCA publications matching "bomba de infusión", 2025-10-15 to 2026-10-15002026-10-15 18:23:39
AEMPSNotas de seguridad de productos sanitarios, 2025-10-15 to 2026-10-15; titles and summaries matched against 10 terms34342026-10-15 18:23:40

Appendix B. Reason codes

CodeMeaning
I1Safety outcome or adverse event
I2Performance or clinical effectiveness
I3Usability or use error
I4State of the art or alternative therapies
X1Not the device type under surveillance
X2Non-human study
X3No safety or performance data (editorial, news)
X4Duplicate of another record
X5Outside the surveillance period
R0Needs a human decision