Prepared 15 October 2026 · Snapshot 2026-10-15-volumetric-infusion-pump-sample-858765
Prepared with Psurio
| Priority | Signal | Evidence |
|---|---|---|
| high | Rising: Pumping Problem | 179 reports in 2026-07-03 to 2026-09-30 (60.54/month), 15.27× the earlier monthly rate. |
| high | Rising: Moisture Damage | 49 reports in 2026-07-03 to 2026-09-30 (16.57/month), 71.05× the earlier monthly rate. |
| high | Rising: Device Displays Incorrect Message | 48 reports in 2026-07-03 to 2026-09-30 (16.23/month), 6.33× the earlier monthly rate. |
| high | Rising: Device Reprocessing Problem | 44 reports in 2026-07-03 to 2026-09-30 (14.88/month), new in the recent window. |
| medium | Rising: Failure to Infuse | 1109 reports in 2026-07-03 to 2026-09-30 (375.09/month), 2.03× the earlier monthly rate. |
| medium | ANSM field safety notice: Pompe à perfusion ambulatoire (accessoire) – Cassette-réservoirs CADD avec raccord NRFit – ICU Medical Inc. | 2026-09-16: Rappel de produit · Rappel n° R2625748 destiné aux pharmacies d'usage intérieur et aux prestataires de santé à domicile (matched device term: pompe à perfusion). |
| medium | MHRA field safety notice: ICU Medical: CADD Medication Cassette Reservoirs with NRFit | 2026-08-28: Infusion pump cassette (matched device term: infusion pump). |
| medium | BfArM field safety notice: Urgent Field Safety Notice for AlarisTM GW 800 Volumetric Pump by BD Switzerland Sarl | 2026-08-05 (matched device term: volumetric pump). |
| medium | ANSM field safety notice: Pompe à perfusion Alaris GW 800 – BD | 2026-08-04: Information aux utilisateurs · Information n° R2622916 destinée aux pharmacies d'usage intérieur, établissements de santé (matched device term: pompe à perfusion). |
| medium | Swissmedic field safety notice: MEDCAPTAIN LIFE SCIENCE CO., LTD.: BD neXus Disposable Infusion Set | 2026-07-22: Electric infusion pump administration set, single-use (update) (matched device term: infusion pump). |
| medium | FDA recall by Baxter Healthcare Corporation (2 products) | Initiated 2026-08-11: A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains. |
| medium | FDA recall by ICU Medical, Inc. (2 products) | Initiated 2026-06-10: In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate. |
| medium | FDA recall by ICU Medical, Inc. | Initiated 2026-05-12: Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and… |
| medium | FDA recall by Fresenius Kabi USA, LLC | Initiated 2026-05-06: Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power. |
| medium | FDA recall by Fresenius Kabi USA, LLC | Initiated 2026-05-06: Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service.… |
| info | 93 reports coded as death | MAUDE reports don't establish that the device caused the death. Read each narrative. |
| info | 16 more field safety notices in the period | Listed in full in the field safety notices section. |
| info | 13 more recall events in the period | Listed in full in the recalls table. |
A signal is a prompt to investigate, not a conclusion. Thresholds: a monthly spike above the mean of earlier months plus two standard deviations; a problem reported at least twice as often per month in the last 90 days, with at least 5 reports; a complaint category significantly above its earlier rate (one-sided Poisson test, p < 0.01, at least 3 complaints).
Hollow points are incomplete months. MAUDE data for this device runs to 2026-09-30. Counts are not adjusted for the number of devices in use.
| Type | Reports |
|---|---|
| Malfunction | 50,321 |
| Injury | 673 |
| Other | 248 |
| Death | 93 |
Recent: 2026-07-03 to 2026-09-30. Earlier: 2025-10-15 to 2026-07-02. Rates are per month. Rows marked “Exact count” were counted one problem at a time; rows marked “Estimate” come from the FDA’s list of the most frequent problems and may undercount less common ones.
| Problem | Recent | Earlier | Recent / month | Earlier / month | Change | Count |
|---|---|---|---|---|---|---|
| Failure to Infuse | 1,109 | 1,582 | 375.1 | 184.5 | 2.03× | Exact count |
| Pumping Problem | 179 | 34 | 60.5 | 4 | 15.27× | Exact count |
| Moisture Damage | 49 | 2 | 16.6 | 0.2 | 71.05× | Exact count |
| Device Displays Incorrect Message | 48 | 22 | 16.2 | 2.6 | 6.33× | Exact count |
| Device Reprocessing Problem | 44 | 0 | 14.9 | 0 | new | Exact count |
| Material Discolored | 44 | 0 | 14.9 | 0 | new | Exact count |
| Leak/Splash | 42 | 35 | 14.2 | 4.1 | 3.48× | Exact count |
| Inaccurate Flow Rate | 34 | 41 | 11.5 | 4.8 | 2.4× | Exact count |
| Defective Device | 24 | 31 | 8.1 | 3.6 | 2.25× | Exact count |
| Intermittent Infusion | 23 | 14 | 7.8 | 1.6 | 4.76× | Exact count |
| Device Alarm System | 2,129 | 3,649 | 720.1 | 425.6 | 1.69× | Estimate |
| Device Sensing Problem | 1,787 | 5,882 | 604.4 | 686 | 0.88× | Estimate |
| Break | 1,326 | 4,125 | 448.5 | 481.1 | 0.93× | Estimate |
| Failure to Sense | 1,315 | 3,875 | 444.8 | 451.9 | 0.98× | Estimate |
| No Apparent Adverse Event | 1,128 | 4,053 | 381.5 | 472.7 | 0.81× | Estimate |
| Patient problem | Reports |
|---|---|
| No Clinical Signs, Symptoms or Conditions | 48660 |
| Insufficient Information | 1110 |
| Hyperglycemia | 944 |
| Low Blood Pressure/ Hypotension | 153 |
| Hypoglycemia | 87 |
| Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available | 85 |
| High Blood Pressure/ Hypertension | 50 |
| Tachycardia | 39 |
| Hemorrhage/Blood Loss/Bleeding | 36 |
| Reaction to Medicinal Component of Device | 32 |
24 recalled products in 18 recall events, newest first.
| Initiated | Firm | Reason | Recall numbers | Status |
|---|---|---|---|---|
| 2026-08-11 | Baxter Healthcare Corporation | A software issue that may occur during a Volume/Time (VOT) multi-syringe infusion when a syringe reaches empty and is replaced while a Volume To Be Infused (VTBI) remains. | Z-3065-2026 Z-3066-2026 | Open, Classified |
| 2026-06-10 | ICU Medical, Inc. | In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate. | Z-2723-2026 Z-2724-2026 | Open, Classified |
| 2026-05-12 | ICU Medical, Inc. | Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall. The presence of discoloration and particulate matter in the fluid path has not been ruled out. | Z-3162-2026 | Open, Classified |
| 2026-05-06 | Fresenius Kabi USA, LLC | Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power. | Z-2223-2026 | Open, Classified |
| 2026-05-06 | Fresenius Kabi USA, LLC | Fresenius Kabi has been made aware via customer complaints that dropping or severe jarring of the Ivenix LVP may cause damage to internal components that may not be apparent or readily visible and the devices therefore remained in service. More specifically, drops or jarring on a hard surface can cause the touchscreen input cable to loosen, resulting in unintended touchscreen behavior on the LVP, including random screen touches and/or lack of screen responsiveness to user input. Under the reported failure condition, unintended touch inputs may register at arbitrary screen locations without user interaction. These inputs occur intermittently and without a predictable pattern. | Z-2224-2026 | Open, Classified |
| 2026-04-29 | ICU Medical, Inc. | Under certain conditions, the pump may not display a soft limit alert, and will not notify the clinician of a soft limit violation before confirming and starting delivery. This may potentially lead to under- or over-delivery. | Z-2311-2026 Z-2312-2026 Z-2313-2026 | Open, Classified |
| 2026-02-11 | CareFusion 303, Inc. | Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units | Z-1645-2026 | Open, Classified |
| 2026-02-04 | Fresenius Kabi USA, LLC | Potential for the Cassette Loading Lever to break. | Z-2274-2026 | Open, Classified |
| 2026-01-13 | Baxter Healthcare Corporation | Pumps were released without full testing being performed, including occlusion alarm testing. | Z-1368-2026 | Open, Classified |
| 2025-12-15 | ICU Medical, Inc. | ICU Medical received one lot of defective primary speakers from our supplier that may fail to produce an audible sound. If the primary speaker in your Plum Duo Infusion Pump fails, the pump may not produce audible alarms, alerts, or provide audible feedback for touchscreen inputs. | Z-1094-2026 | Open, Classified |
| 2025-12-15 | ICU Medical, Inc. | Plum Solo and Duo Infusion pumps include a feature, that when selected, automatically flushes the downstream line after a piggyback therapy. The programmed flush volume is delivered from the primary line container at the piggyback therapy rate after the piggyback therapy is complete. In certain cases, as described below, the clinician may receive an Upper Hard Limit (UHL) or Lower Hard Limit (LHL) Violation message when programming a piggyback flush, which prevents the flush from being programmed. | Z-1118-2026 Z-1119-2026 Z-1120-2026 | Open, Classified |
| 2025-11-28 | Baxter Healthcare Corporation | Certain pumps have potentially been released from service with defective grease applied to the cam and motor gears. Due to its low viscosity, the defective grease may lead to the device having insufficient lubrication on the cam, which could lead to premature wear of the mechanism assembly, resulting in over-infusion or a free-flow situation. | Z-1121-2026 | Open, Classified |
| 2025-11-21 | Fresenius Kabi USA, LLC | Emphasizing instructions for LVP duration programming located in the IFU. | Z-1048-2026 | Open, Classified |
| 2025-11-14 | Fresenius Kabi USA, LLC | Software version 5.10.1 and earlier contain anomalies that have the potential to cause serious patient harm or death. | Z-0885-2026 | Open, Classified |
| 2025-11-06 | CareFusion 303, Inc. | Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital network | Z-1217-2026 | Open, Classified |
| 2025-11-03 | Fresenius Kabi USA, LLC | Incorrect assembly of specific lot of LVP Primary Administration Set that could result in serious patient injury. | Z-0586-2026 | Open, Classified |
| 2025-11-03 | Fresenius Kabi USA, LLC | Downstream Occlusion alarms may occur during procedures using low flow infusion rates when a check valve* is added to the Ivenix Administration Set. | Z-0848-2026 | Open, Classified |
| 2025-10-17 | CareFusion 303, Inc. | If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated. | Z-0430-2026 | Open, Classified |
34 notices from European regulators (MHRA, BfArM, ANSM, Swissmedic, AEMPS) in the period, of which 13 only mention a search term inside the notice document.
| Date | Source | Notice | Matched on |
|---|---|---|---|
| 2026-09-16 | ANSM Field safety notice | Pompe à perfusion ambulatoire (accessoire) – Cassette-réservoirs CADD avec raccord NRFit – ICU Medical Inc. Rappel de produit · Rappel n° R2625748 destiné aux pharmacies d'usage intérieur et aux prestataires de santé à domicile | device term: pompe à perfusion |
| 2026-08-28 | MHRA Field safety notice | ICU Medical: CADD Medication Cassette Reservoirs with NRFit Infusion pump cassette Ref 40630462 2026/008/025/601/011 | device term: infusion pump |
| 2026-08-05 | BfArM Field safety notice | Urgent Field Safety Notice for AlarisTM GW 800 Volumetric Pump by BD Switzerland Sarl | device term: volumetric pump |
| 2026-08-04 | ANSM Field safety notice | Pompe à perfusion Alaris GW 800 – BD Information aux utilisateurs · Information n° R2622916 destinée aux pharmacies d'usage intérieur, établissements de santé | device term: pompe à perfusion |
| 2026-07-22 | Swissmedic Field safety notice | MEDCAPTAIN LIFE SCIENCE CO., LTD.: BD neXus Disposable Infusion Set Electric infusion pump administration set, single-use (update) Ref Vk_20260713_032 | device term: infusion pump |
| 2026-07-02 | Swissmedic Field safety notice | B. Braun Melsungen AG: Infusomat Space P Bedside infusion pump, single-channel Ref Vk_20260701_035 | device term: infusion pump |
| 2026-03-31 | MHRA Field safety notice | Abiomed, Inc: Impella Purge Cassette Infusion Pump Cassette · Model: 0043-0001;0043-0009 Ref 39420703 2026/004/002/601/034 | device term: infusion pump |
| 2026-02-16 | ANSM Field safety notice | Pompe à perfusion ambulatoire (accessoires) – Set d’administration CADD à débit élevé – ICU Medical Inc. Rappel de produit · Rappel n° R2604556 destiné aux pharmacies d'usage intérieur des établissements de santé et aux prestataires de santé à domicile | device term: pompe à perfusion |
| 2026-02-13 | MHRA Field safety notice | ICU Medical: CADD High Volume Administration Sets Electric infusion pump administration set · Model: 21-7361-24 Ref 38755259 2026/002/010/601/028 | device term: infusion pump |
| 2026-02-12 | Swissmedic Field safety notice | ICU Medical Inc: CADD™ High Volume Administration Sets Electric infusion pump administration set, single-use Ref Vk_20260210_023 | device term: infusion pump |
| 2026-02-05 | ANSM Field safety notice | Pompe à perfusion (tubulure) – Sets de perfusion Alaris – BD Rappel de produit · Information n° R2602319 destinée aux pharmacies d'usage intérieur, établissements de santé | device term: pompe à perfusion |
| 2026-01-28 | MHRA Field safety notice | Baxter: EVO IQ Large Volume Pump Model: ELVP002UKI and 04400121200 Ref 38329736 2026/001/019/601/021 | device term: large volume pump |
| 2026-01-27 | Swissmedic Field safety notice | BD Switzerland Sàrl: Alaris VP Infusion Sets Electric infusion pump administration set, single-use Ref Vk_20260122_01 | device term: infusion pump |
| 2026-01-21 | MHRA Field safety notice | BD: Alaris VP Infusion Sets Infusion Pump Administration Set Ref 38309876 2026/001/014/601/056 | device term: infusion pump |
| 2025-12-17 | ANSM Field safety notice | Pompe à perfusion BD neXus V700 et pousse-seringue – BD neXus S700 – Medcaptain Life Science Co. Information aux utilisateurs · Information n° R2534072 destinée aux pharmacies d'usage intérieur des établissements de santé | device term: pompe à perfusion |
| 2025-12-16 | BfArM Field safety notice | Urgent Field Safety Notice for BD Alaris System Infusion Pump by BD Switzerland Sarl | device term: infusion pump |
| 2025-12-15 | MHRA Field safety notice | MEDCAPTAIN MEDICAL TECHNOLOGY: BD neXus V700 Infusion Pump & S700 Syringe Pump V700: Infusion Pump; S700: Syringe Pump Ref 37923637 2025/012/008/601/002 | device term: infusion pump |
| 2025-12-01 | MHRA Field safety notice | Baxter: Ambulatory Elastomeric Infusion Pumps Ref 37850863 2025/012/001/601/064 | device term: infusion pump |
| 2025-12-01 | BfArM Field safety notice | Urgent Field Safety Notice for Ambulatory Elastomeric Infusion Pumps by Baxter Healthcare S. (Opfikon) | device term: infusion pump |
| 2025-10-28 | MHRA Field safety notice | Baxter Healthcare SA: EVO IQ Large Volume Pump 15th August 2025 Bedside infusion pump, single-channel MHRA reference: 37397515 2024/011/028/601/018 – Customer reply form · Model: ELVP001UKI Ref 37397515 2024/011/028/601/018 | device term: infusion pump |
| 2025-10-16 | MHRA Field safety notice | ICU Medical: CADD-Solis Ambulatory Infusion Pumps Bedside infusion pump, single-channel · Model: CADD-Solis VIP Ref 37337852 2025/010/013/601/015 | device term: infusion pump |
BfArM's search matched these in the notice text. Check whether each concerns the device.
Records were ordered for review by a transparent relevance score (device in title or subject indexing, mentions in the abstract, safety topics). The score only ordered the work; every record was decided individually.
| Band | Records | Included | Excluded | Awaiting |
|---|---|---|---|---|
| high | 26 | 0 | 0 | 26 |
| medium | 15 | 0 | 0 | 15 |
| low | 127 | 0 | 88 | 39 |
| Code | Reason | Records |
|---|---|---|
| X3 | No safety or performance data (editorial, news) | 1 |
| X1 | Not the device type under surveillance | 87 |
No records included yet. Review the queue to complete this section.
| Received | Report |
|---|---|
| 2026-09-30 Malfunction | Spectrum IQ BAXTER HEALTHCARE CORPORATION · 1314492-2026-02104 False Alarm |
| 2026-09-30 Malfunction | Alaris System CAREFUSION 303, INC. · 2016493-2026-67346 Break; Contamination It was reported that the device was missing a piece. There was no patient involvement. |
| 2026-09-30 Malfunction | Alaris System CAREFUSION 303, INC. · 2016493-2026-67349 Circuit Failure; Physical Resistance/Sticking It was reported that the device had pump will not power on. There was no patient involvement. |
| 2026-09-30 Malfunction | Alaris System CAREFUSION 303, INC. · 2016493-2026-67350 Contamination It was reported that the device had not turning on, replaced battery, same issue. There was no patient involvement. |
| 2026-09-30 Malfunction | Alaris System CAREFUSION 303, INC. · 2016493-2026-67354 Device Sensing Problem It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement. |
| 2026-09-30 Malfunction | Alaris System CAREFUSION 303, INC. · 2016493-2026-67353 Calibration Problem It was reported that a failure was observed during a planned preventative maintenance, recall remediation, or repair order service event. "[Field Service]";"[Door ajar-safety clamp test failed]". There was no reported patient involvement. |
| 2026-09-30 Malfunction | Alaris System CAREFUSION 303, INC. · 2016493-2026-67347 Device Alarm System It was reported that the device had error code 242.4030. There was no patient involvement. |
| 2026-09-30 Malfunction | Alaris System CAREFUSION 303, INC. · 2016493-2026-67351 Circuit Failure It was reported that the device washed out screen. There was patient involvement but unknown harm. |
| 2026-09-30 Malfunction | Alaris System CAREFUSION 303, INC. · 2016493-2026-67348 Failure to Sense It was reported that the device displayed channel error code 242.4030. There was no patient involvement. |
| 2026-09-30 Malfunction | Alaris System CAREFUSION 303, INC. · 2016493-2026-67355 Premature Discharge of Battery It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement. |
| 2026-09-30 Malfunction | IVENIX INFUSION SYSTEM FRESENIUS KABI USA, LLC · 3014732157-2026-04133 No Apparent Adverse Event The following has been reported:The unit was returned during service after initial complaint investigation was completed for rear housing issues. Secondary Findings: Unit experienced tubing set misloaded warnings due to sticky FVA pins. Pins and channels were cleaned and unit then passed a mock infusion. |
| 2026-09-30 Malfunction | IVENIX INFUSION SYSTEM FRESENIUS KABI USA, LLC · 3014732157-2026-04134 No Apparent Adverse Event The following has been reported:The unit was returned during service after initial complaint investigation was completed for lever arm issues. Secondary Findings: Secondary Findings: Outlet FVA pin stuck in due to excess debris. FVA pin lip seals will be replaced during repair. |
| Source | Query | Hits | Retrieved | Run at (UTC) |
|---|---|---|---|---|
| PubMed | ("infusion pump"[tiab] OR "volumetric pump"[tiab] OR "smart pump"[tiab] OR "large volume pump"[tiab] OR "Infusion Pumps"[mh]) AND ("adverse event"[tiab] OR "malfunction"[tiab] OR "device failure"[tiab] OR "equipment failure"[tiab] OR "recall"[tiab] OR "occlusion"[tiab] OR "alarm"[tiab] OR "free flow"[tiab] OR "over-infusion"[tiab] OR "under-infusion"[tiab] OR "medication error"[tiab] OR "programming error"[tiab] OR "dose error"[tiab] OR "drug library"[tiab] OR "use error"[tiab] OR "interoperability"[tiab] OR "Equipment Failure"[mh] OR "Equipment Safety"[mh] OR "Medication Errors"[mh] OR "Equipment Failure Analysis"[mh] OR "adverse effects"[sh]) | publication date 2025/10/15 to 2026/10/15 | 167 | 167 | 2026-10-15 18:22:46 |
| Europe PMC | (TITLE_ABS:"infusion pump" OR TITLE_ABS:"volumetric pump" OR TITLE_ABS:"smart pump" OR TITLE_ABS:"large volume pump") AND (TITLE_ABS:"adverse event" OR TITLE_ABS:"malfunction" OR TITLE_ABS:"device failure" OR TITLE_ABS:"equipment failure" OR TITLE_ABS:"recall" OR TITLE_ABS:"occlusion" OR TITLE_ABS:"alarm" OR TITLE_ABS:"free flow" OR TITLE_ABS:"over-infusion" OR TITLE_ABS:"under-infusion" OR TITLE_ABS:"medication error" OR TITLE_ABS:"programming error" OR TITLE_ABS:"dose error" OR TITLE_ABS:"drug library" OR TITLE_ABS:"use error" OR TITLE_ABS:"interoperability") AND FIRST_PDATE:[2025-10-15 TO 2026-10-15] AND (SRC:MED OR SRC:PMC OR SRC:PPR) | 23 | 23 | 2026-10-15 18:22:46 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND date_received:[20251015 TO 20261015] | count by date_received | 51335 | 312 | 2026-10-15 18:22:47 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND date_received:[20260703 TO 20260930] | count by product_problems.exact | 16623 | 100 | 2026-10-15 18:22:47 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND date_received:[20251015 TO 20260702] | count by product_problems.exact | 49394 | 100 | 2026-10-15 18:22:47 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Failure to Infuse" AND date_received:[20260703 TO 20260930] | 1109 | 1 | 2026-10-15 18:22:48 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Failure to Infuse" AND date_received:[20251015 TO 20260702] | 1582 | 1 | 2026-10-15 18:22:48 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Pumping Problem" AND date_received:[20260703 TO 20260930] | 179 | 1 | 2026-10-15 18:22:49 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Pumping Problem" AND date_received:[20251015 TO 20260702] | 34 | 1 | 2026-10-15 18:22:49 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Moisture Damage" AND date_received:[20260703 TO 20260930] | 49 | 1 | 2026-10-15 18:22:49 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Moisture Damage" AND date_received:[20251015 TO 20260702] | 2 | 1 | 2026-10-15 18:22:49 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Device Displays Incorrect Message" AND date_received:[20260703 TO 20260930] | 48 | 1 | 2026-10-15 18:22:50 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Device Displays Incorrect Message" AND date_received:[20251015 TO 20260702] | 22 | 1 | 2026-10-15 18:22:50 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Device Reprocessing Problem" AND date_received:[20260703 TO 20260930] | 44 | 1 | 2026-10-15 18:22:50 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Device Reprocessing Problem" AND date_received:[20251015 TO 20260702] | 0 | 0 | 2026-10-15 18:22:51 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Material Discolored" AND date_received:[20260703 TO 20260930] | 44 | 1 | 2026-10-15 18:22:51 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Material Discolored" AND date_received:[20251015 TO 20260702] | 0 | 0 | 2026-10-15 18:22:51 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Leak/Splash" AND date_received:[20260703 TO 20260930] | 42 | 1 | 2026-10-15 18:22:52 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Leak/Splash" AND date_received:[20251015 TO 20260702] | 35 | 1 | 2026-10-15 18:22:52 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Inaccurate Flow Rate" AND date_received:[20260703 TO 20260930] | 34 | 1 | 2026-10-15 18:22:52 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Inaccurate Flow Rate" AND date_received:[20251015 TO 20260702] | 41 | 1 | 2026-10-15 18:22:53 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Defective Device" AND date_received:[20260703 TO 20260930] | 24 | 1 | 2026-10-15 18:22:53 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Defective Device" AND date_received:[20251015 TO 20260702] | 31 | 1 | 2026-10-15 18:22:53 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Intermittent Infusion" AND date_received:[20260703 TO 20260930] | 23 | 1 | 2026-10-15 18:22:53 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND product_problems.exact:"Intermittent Infusion" AND date_received:[20251015 TO 20260702] | 14 | 1 | 2026-10-15 18:22:54 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND date_received:[20251015 TO 20261015] | count by event_type.exact | 51335 | 4 | 2026-10-15 18:22:54 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND date_received:[20251015 TO 20261015] | count by patient.patient_problems.exact | 51500 | 25 | 2026-10-15 18:22:54 |
| openFDA MAUDE | (device.device_report_product_code:"FRN") AND date_received:[20251015 TO 20261015] | sorted by date_received:desc | 51335 | 100 | 2026-10-15 18:22:55 |
| openFDA recalls | (product_code:"FRN") AND event_date_initiated:[2025-10-15 TO 2026-10-15] | 24 | 24 | 2026-10-15 18:22:57 |
| MHRA | GOV.UK medical safety alerts of type field-safety-notices (and titled "Field Safety Notices"), published 2025-10-15 to 2026-10-15; each notice matched against: infusion pump, volumetric pump, smart pump, large volume pump, pompe à perfusion, pompes à perfusion, Infusionspumpe, pompa infusionale, pompa per infusione, bomba de infusión | 553 | 9 | 2026-10-15 18:23:04 |
| MHRA | GOV.UK medical safety alerts of type device-safety-information, published 2025-10-15 to 2026-10-15; title and summary matched against the same terms | 14 | 0 | 2026-10-15 18:23:04 |
| BfArM | Field safety notices (Kundeninfo) containing "infusion pump", newest first, dated 2025-10-15 to 2026-10-15 | 14 | 14 | 2026-10-15 18:23:06 |
| BfArM | Field safety notices (Kundeninfo) containing "volumetric pump", newest first, dated 2025-10-15 to 2026-10-15 | 3 | 3 | 2026-10-15 18:23:08 |
| BfArM | Field safety notices (Kundeninfo) containing "smart pump", newest first, dated 2025-10-15 to 2026-10-15 | 4 | 4 | 2026-10-15 18:23:10 |
| BfArM | Field safety notices (Kundeninfo) containing "large volume pump", newest first, dated 2025-10-15 to 2026-10-15 | 0 | 0 | 2026-10-15 18:23:13 |
| BfArM | Field safety notices (Kundeninfo) containing "pompe à perfusion", newest first, dated 2025-10-15 to 2026-10-15 | 0 | 0 | 2026-10-15 18:23:14 |
| BfArM | Field safety notices (Kundeninfo) containing "pompes à perfusion", newest first, dated 2025-10-15 to 2026-10-15 | 0 | 0 | 2026-10-15 18:23:16 |
| BfArM | Field safety notices (Kundeninfo) containing "Infusionspumpe", newest first, dated 2025-10-15 to 2026-10-15 | 0 | 0 | 2026-10-15 18:23:17 |
| BfArM | Field safety notices (Kundeninfo) containing "pompa infusionale", newest first, dated 2025-10-15 to 2026-10-15 | 0 | 0 | 2026-10-15 18:23:19 |
| BfArM | Field safety notices (Kundeninfo) containing "pompa per infusione", newest first, dated 2025-10-15 to 2026-10-15 | 0 | 0 | 2026-10-15 18:23:20 |
| BfArM | Field safety notices (Kundeninfo) containing "bomba de infusión", newest first, dated 2025-10-15 to 2026-10-15 | 0 | 0 | 2026-10-15 18:23:21 |
| ANSM | Informations de sécurité, dispositifs médicaux, 2025-10-15 to 2026-10-15; titles matched against 10 terms | 569 | 569 | 2026-10-15 18:23:36 |
| Swissmedic | FSCA publications matching "infusion pump", 2025-10-15 to 2026-10-15 | 15 | 15 | 2026-10-15 18:23:36 |
| Swissmedic | FSCA publications matching "volumetric pump", 2025-10-15 to 2026-10-15 | 1 | 1 | 2026-10-15 18:23:36 |
| Swissmedic | FSCA publications matching "smart pump", 2025-10-15 to 2026-10-15 | 2 | 2 | 2026-10-15 18:23:37 |
| Swissmedic | FSCA publications matching "large volume pump", 2025-10-15 to 2026-10-15 | 1 | 1 | 2026-10-15 18:23:37 |
| Swissmedic | FSCA publications matching "pompe à perfusion", 2025-10-15 to 2026-10-15 | 8 | 8 | 2026-10-15 18:23:37 |
| Swissmedic | FSCA publications matching "pompes à perfusion", 2025-10-15 to 2026-10-15 | 4 | 4 | 2026-10-15 18:23:38 |
| Swissmedic | FSCA publications matching "Infusionspumpe", 2025-10-15 to 2026-10-15 | 3 | 3 | 2026-10-15 18:23:38 |
| Swissmedic | FSCA publications matching "pompa infusionale", 2025-10-15 to 2026-10-15 | 1 | 1 | 2026-10-15 18:23:39 |
| Swissmedic | FSCA publications matching "pompa per infusione", 2025-10-15 to 2026-10-15 | 5 | 5 | 2026-10-15 18:23:39 |
| Swissmedic | FSCA publications matching "bomba de infusión", 2025-10-15 to 2026-10-15 | 0 | 0 | 2026-10-15 18:23:39 |
| AEMPS | Notas de seguridad de productos sanitarios, 2025-10-15 to 2026-10-15; titles and summaries matched against 10 terms | 34 | 34 | 2026-10-15 18:23:40 |
| Code | Meaning |
|---|---|
| I1 | Safety outcome or adverse event |
| I2 | Performance or clinical effectiveness |
| I3 | Usability or use error |
| I4 | State of the art or alternative therapies |
| X1 | Not the device type under surveillance |
| X2 | Non-human study |
| X3 | No safety or performance data (editorial, news) |
| X4 | Duplicate of another record |
| X5 | Outside the surveillance period |
| R0 | Needs a human decision |