The EU Medical Device Regulation (MDR)

Regulation (EU) 2017/745, the MDR, sets the rules for placing medical devices on the EU market: how they’re classified and certified, and what manufacturers must do once they’re on sale. It has applied since 26 May 2021, replacing the Medical Devices Directive and the Active Implantable Medical Devices Directive.

Device classes

Devices are classed by risk using the 22 rules in Annex VIII, based on things like how long they’re in contact with the body, how invasive they are, and whether they’re active. The class decides how the device is certified and how much post-market reporting it needs.

EU MDR device classes, certification and post-market reports
ClassExamplesCertificationPost-market report
Class IBandages, non-invasive instrumentsSelf-declared; a notified body for sterile (Is), measuring (Im) and reusable surgical (Ir) aspectsPMS report
Class IIaHearing aids, short-term cathetersNotified bodyPSUR at least every 2 years
Class IIbInfusion pumps, ventilatorsNotified bodyPSUR at least every year
Class IIIHeart valves, implants with medicinesNotified bodyPSUR every year, evaluated by the notified body

Who does what

  • Manufacturer (Article 10): responsible for the device, its quality management system, risk management, clinical evaluation, technical documentation and post-market surveillance.
  • PRRC (Article 15): at least one person responsible for regulatory compliance, with set qualifications.
  • Authorised representative (Article 11): required for manufacturers outside the EU.
  • Importers and distributors (Articles 13 and 14): check the device and its documentation, and pass on complaints and incidents.
  • Notified bodies: independent organisations designated to assess conformity and audit manufacturers.
  • Competent authorities: each member state’s regulator, which receives vigilance reports and carries out market surveillance.

What the MDR expects after launch

The MDR puts much more weight on the device’s life after launch than the directives did. Manufacturers must run post-market surveillance under a written plan, report serious incidents and field safety corrective actions within set deadlines (Articles 87 to 89), report significant trends (Article 88), keep the clinical evaluation up to date with post-market clinical follow-up (Annex XIV, Part B), and write a PSUR or a PMS report.

Devices and economic operators are registered in EUDAMED, the European database, and devices carry a unique device identifier (UDI) so they can be traced.

Transition deadlines

Regulation (EU) 2023/607 extended the time devices certified under the old directives (“legacy devices”) can stay on the market, provided the manufacturer had a quality management system in place and applied to a notified body by 26 May 2024, and signed a written agreement with it by 26 September 2024:

  • 31 December 2027: class III devices, and class IIb implantable devices except some simple ones such as sutures, staples, screws and plates.
  • 31 December 2028: other class IIb devices, class IIa devices, and class I devices that are sterile or have a measuring function.

Legacy devices still have to meet the MDR’s post-market surveillance, vigilance and registration requirements during the transition (Article 120(3)).

The UK

Great Britain has its own rules, the UK Medical Devices Regulations 2002. Since 16 June 2025 they include post-market surveillance requirements close to the MDR’s: a PMS plan, PSURs for class IIa and above, a PMS report for class I every 3 years, and serious incidents reported within 15 days. Northern Ireland follows the EU MDR. Great Britain also accepts CE marked devices for a transition period; check the MHRA’s current dates.

How Psurio helps

In the product

Psurio covers the post-market side of the MDR for each device: the surveillance data, signals, literature screening, risk file links, the PSUR or PMS report, and a second person’s sign-off, all on an audit trail. See how it works.

Common questions

Does the MDR apply to manufacturers outside the EU?
Yes, to any device placed on the EU market. A manufacturer outside the EU must appoint an EU authorised representative (Article 11), who is named on the label and shares responsibility for compliance.
What is a PRRC?
The person responsible for regulatory compliance (Article 15): a qualified person who makes sure, among other things, that post-market surveillance and vigilance obligations are met. Micro and small enterprises may use an external PRRC who is permanently and continuously available to them.
Does the UK follow the EU MDR?
Northern Ireland follows the EU MDR. Great Britain has its own rules, the UK Medical Devices Regulations 2002, which since June 2025 include post-market surveillance requirements close to the MDR's. Great Britain also accepts CE marked devices for a transition period.