Skip to content

Regulations · PSUR

The periodic safety update report (PSUR)

A PSUR is the regular, signed summary of what post-market surveillance found for a device of class IIa or above, and whether its benefits still outweigh its risks. It’s required by Article 86 of the EU MDR and, in Great Britain, by the UK rules in force since June 2025.

What a PSUR is

Every manufacturer runs a post-market surveillance system that keeps collecting data on its devices once they’re on the market: complaints, incidents, published studies, registries, and similar devices’ problems (see post-market surveillance). The PSUR is where that data is analysed and concluded on, at fixed intervals, for devices of class IIa, IIb and III.

It is part of the device’s technical documentation on post-market surveillance (Annex III of the MDR), so it is kept with the rest of the file and is one of the first things a notified body reads.

What it must contain

Article 86(1) sets the minimum. A PSUR summarises the results and conclusions of the analysis of post-market surveillance data, with the rationale and description of any preventive and corrective actions taken, and states:

  • The conclusions of the benefit-risk determination: whether the device’s benefits still outweigh its risks.
  • The main findings of post-market clinical follow-up (PMCF), the clinical data gathered after launch.
  • The volume of sales, and an estimate of the size and characteristics of the population using the device and, where practicable, how often it is used.

MDCG 2022-21, the Medical Device Coordination Group’s guidance on PSURs, turns this into a standard structure: device and certificate details, the data collection period, sales and population, serious incidents and field safety corrective actions, non-serious incidents and trends, feedback and complaints, literature and similar devices, PMCF, and the overall conclusions.

How often, and who reviews it

PSUR frequency and review by device class under the EU MDR
ClassHow oftenWhat happens to it
Class III and implantable devicesAt least every yearSubmitted to the notified body, which evaluates it and adds its evaluation
Class IIb (not implantable)At least every yearKept in the technical documentation and made available to the notified body
Class IIaWhen necessary, and at least every 2 yearsKept in the technical documentation and made available to the notified body
Class INo PSUR: a PMS report insteadSee the PMS report guide

For class III and implantable devices, Article 86(2) says the PSUR is submitted through EUDAMED, the EU database. Until EUDAMED’s vigilance and post-market surveillance module is in use, agree with your notified body how to send it. Competent authorities can ask for any PSUR at any time.

In Great Britain

Since 16 June 2025, Part 4A of the UK Medical Devices Regulations 2002 (added by SI 2024/1368) has its own PSUR rule, regulation 44ZM. It applies to devices placed on the market in Great Britain, whether UKCA or CE marked:

  • Most devices needing a PSUR: the first within a year of being placed on the market, then at least every year.
  • Class IIa: the first within 2 years, then at least every 2 years.
  • Submission: to the UK approved body for the device, if there is one, which reports back on class III and implantable devices. The MHRA can ask for a PSUR at any time and must get it within 3 working days.

Where PSURs go wrong

  • Literature and database searches that can't be repeated, because the exact queries and dates weren't recorded.
  • Complaint trends described in words, with no stated method or threshold for what counts as a significant rise.
  • Findings that never reach the risk file: new hazards, or harms happening more often than the risk analysis estimated.
  • Sales and population figures missing or unexplained, so rates can't be calculated.
  • A benefit-risk conclusion that restates the last one instead of following from this period's data.

How Psurio helps

In the product

A snapshot collects the period’s literature, FDA adverse events and recalls, and European field safety notices for the device, logs every query, and flags signals. Imported complaints are trended with a stated statistical test, and findings are linked to hazards in the risk file.

The PSUR draft follows the MDCG 2022-21 structure, you add sales, population, PMCF and the benefit-risk conclusion, and a second person signs it off with the content fingerprinted. See how it works.

Common questions

Do class I devices need a PSUR?
No. Under the EU MDR, class I devices need a post-market surveillance report (Article 85) instead, updated when necessary. In Great Britain they need a post-market surveillance report at least every 3 years.
Can one PSUR cover several devices?
Yes. Article 86 asks for a PSUR for each device and, where relevant, for each category or group of devices. MDCG 2022-21 explains when devices can be grouped and how to pick the leading device.
Is the MDCG 2022-21 template mandatory?
MDCG guidance isn't legally binding, but notified bodies expect PSURs to follow its structure, so departing from it needs a good reason.
Does a PSUR replace vigilance reporting?
No. Serious incidents and field safety corrective actions are reported on their own deadlines (Articles 87 and 89). The PSUR summarises them over the period, alongside everything else the post-market surveillance plan collected.

Sources

Last reviewed 15 October 2026. This guide is general information, not regulatory or legal advice: the legal texts and your notified body or approved body have the final word.

Turn the requirements into a signed report.

Try it free for 14 days, with up to 5 devices. Run a snapshot, screen the literature, and see what a signed report looks like.